Storage temperature range – +5 °C to + 55°C
Operating temperature range – +10 °C to + 35
Specimen feed – Horizontal approx. 30mm/1.18 inches, Vertical Approx. 70mm /2.76 inches
Coarse feed wheel must be closer to the user for fatigue free sectioning.
Adjustable hand-wheel balance system to protect the specimen (object head must not drop in to knife when sectioning is stopped)
Precise orientation system: 8° x/y precision orientation system with calibrated controls for quick orientation of specimen for re-cut. System must have provision to rapidly return to exact zero position which is marked by colored indicators.
Hand wheel brake to allow locking of hand wheel in upper position and locking system below to lock in any position.
Section thickness feed to the specimen from 0.5-60 µm with increments of
0-2 µm in 0.5 µm steps
2 – 10 µm in 1 µm steps
10 – 20 µm in 2 µm steps
20-60 µm in 5 µm steps
Relative humidity: maximum 80% non-condensing.
Storage humidity < 80%.
Section thickness range – 1.0 – 60.0 µm
Specimen retraction must be turned off manually approximately in range of 35-50 µm
Vertical stroke: 70-75 mms Maximum cutting range without retraction 69 mm (without specimen orientation 1 µm)
Equipment must have option of horizontal and vertical orientation (Horizontal angle 8°and vertical angle 8°.
Angle of rotation must be +90°.
Trimming stages option of 10 µm and 50 µm.
Mechanical trimming function with trimming lever having 3 notching positions (0, 10 and 30 µm)
Repositioning of knife holder base North – South : +25 mm
Maximum specimen size 4-5 cm / 2 inches.
Blade holder for both high profile & low profile with proper base
Universal specimen holder is provided
Knife holder must not be available in the instrument